Medición de la densidad y la concentración
Bs. 0.00Bs. 0.00
Medición de la densidad y la concentración
Puntos de control
| VariDens C | VariDens B | |
| Measuring range | 0 – 3 g/cm3 | |
| Temperature range | 10 – 95°C* (regulated Peltier thermoelement) | |
| Pressure range | 0 to 10 Bar (0 to 145 psi) | |
| Density accuracy | ± 0.0001 g/cm3 | ± 0.00005 g/cm3 |
| Reproducibility | ± 0.0001 g/cm3 | ± 0.00002 g/cm3 |
| Resolution | 0.0001 g/cm3 | 0.00001 g/cm3 |
| Temperature accuracy | ± 0.05 °C | ± 0.03 °C |
| Reproducibility | ± 0.02 °C | ± 0.01 °C |
| Resolution | 0.01 °C | 0.01 °C |
| Temperature stabilization time | 25 °C to 20 °C, ± 0.01 °C in 18 minutes* | |
| Wetted parts | Borosilicate glass, PTFE | |
| Bubble detection | Yes (combined: visual-electronic) | |
| Drying | Integrated air pump | |
| Temperature compensation | Yes | |
| Minimum sample volume | 1.5 ml | |
| Measuring time per sample** | < 10 sec | < 30 sec |
| Display | 7″ touchscreen***, 800 x 480 Pixel | |
| Interfaces | Ethernet 10/100/1000, RS232, USB, WLAN*** | |
| Ambient conditions | Temperature 10 °C to 35 °C, Humidity < 85 % not condensing | |
| Power | VAC 90-265; 50/60 Hz; 150 W max | |
| Conformity | International Pharmacopoeia, ASTM, AOAC, DIN, 21 CFR part 11, ICUMSA and others | |
| Dimensions | With display: 430 x 300 x 235 mm (w x d x h) / without display: 430 x 300 x 150 mm (w x d x h) | |
| Highlights | Modularity; user friendliness; fast temperature control; easy cleaning; virtually maintenance free product, 21 CFR part 11 conform; ELN; LIMS ready |
|
* Valid for ambient temp. = 20 °C
** After temperature stabilisation
*** Optional