WaterGiene ATP Swab – for Water Quality

Bs. 0.00Bs. 0.00

WaterGiene ATP Swab – for Water Quality

The WaterGiene® Swab by Charm® is a simple, 20-second test with enhanced sensitivity that measures total ATP (adenosine triphosphate) in water samples and on moist surfaces.

The presence of ATP is a reliable indicator of biological contamination, as it is associated with microorganisms as well as plant, animal, food, and other organic residues.

Thanks to its high sensitivity, WaterGiene® enables rapid corrective action in critical environments such as:

  • Purified water systems

  • Food processing facilities

  • Healthcare and hospital environments

WaterGiene™ – ATP Swab for Water Quality Control

Fast and accurate ATP detection in water and moist surfaces, with results in just 20 seconds.

  • High sensitivity: Detects low levels of biological contamination (microorganisms, organic residues).

  • Versatile application: Use directly in water samples by shaking or on surfaces by swabbing.

  • Recyclable components: Supports sustainability and waste reduction programs.

  • Ideal for rapid hygiene verification in food processing plants, purified water systems, hospitals, and other critical sanitary environments.

  • Enables immediate corrective action, reducing the risk of cross-contamination or cleaning failures.

  • Compatible with the novaLUM II-X luminometer and novaLINK software for data management and audit compliance.

WaterGiene Procedural Flow Chart

paso 1.3 | WaterGiene ATP Swab - for Water Quality Withdraw the swab by gently
pulling and twisting the swab

handle.
paso 2.3 | WaterGiene ATP Swab - for Water Quality Saturate swab by dipping into
liquid for 5 seconds.
paso 3.3 | WaterGiene ATP Swab - for Water Quality Reinsert the swab and twist the
handle down. Gently shake side to

side 3 times.
paso 4.3 | WaterGiene ATP Swab - for Water Quality Unscrew swab handle completely
and shake WaterGiene test 3 times

while holding handle so that liquid

is forced down into the microtube.

Twist handle down again.
paso 5 | WaterGiene ATP Swab - for Water Quality Remove black microcap and verify
the white tablet is broken up

before inserting into the novaLUM

II-X system.

Sensitivity WaterGiene®

Bacterial§ Concentration (cfu/mL WaterGiene Swab mean RLU
% positive results
2,100,000 cfu/mL
210,000 cfu/mL 1,847,967 100%
21,000 cfu/mL 73,581 100%
2,100 cfu/mL 0 0%
150 cfu/mL 0 0%
  • § Four different bacteria studied, E. coli, Pseudomonas, Citrobacter, and Enterobacter. Bacteria were cultured then rinsed in saline and added to pure water
  • Divide by 25 to convert to approximate surface concentration, cfu/100cm²

 

Description

WaterGiene ATP Swab – for Water Quality

The Charm® WaterGiene® ATP Swab (WaterGiene swab) is a simple 20-second test with enhanced sensitivity measuring total ATP (adenosine triphosphate) in water and on wet surfaces. ATP presence is an indicator of microorganisms, plant, animal, or food material, and other biologicals.
The sensitive WaterGiene swab provides fast remediation for purified water systems, food
processing plants, and hospitals.

Application

Scope of application Specific example
Clean In Place (CIP) Systems Sistemas automáticos de limpieza para equipos y tuberías sin necesidad de desmontaje.
Cooling Water Agua utilizada para enfriar sistemas y equipos en procesos industriales.
Animal Laboratory Housing Infraestructura para el alojamiento de animales en laboratorios de investigación.
Fermentation/Tablet Press Equipment Equipos utilizados para la fermentación o la fabricación de tabletas en la industria farmacéutica.
Purified Water Systems Sistemas diseñados para la purificación del agua, asegurando su calidad para uso en laboratorios.
Clean Rooms Salas con control estricto de condiciones ambientales, especialmente para la fabricación y pruebas en ambientes estériles.
In Line Filters Filtros instalados directamente en las líneas de proceso para eliminar contaminantes.
Rinse Water Agua utilizada en procesos de enjuague, esencial para evitar la contaminación cruzada.
Sterile/Non-Sterile Vial & Syringe Filling Equipos para el llenado de viales y jeringas, tanto estériles como no estériles.
Storage Tanks/Silos Tanques y silos para almacenamiento de líquidos o sólidos en diferentes industrias.
Tap and Supply Water Agua proveniente de la red pública o fuentes específicas para diversos procesos.
Surgical Instruments Instrumental quirúrgico utilizado en procedimientos médicos.